Nutrient dosing · PR-FLD-001

Fluidic safety and maintenance

Interlocks, priming, purge, leak control, stock identification, tube replacement and cleaning for the dosing subsystem.

MarkdownJSONRevised 2026-07-26
In brief

Interlocks, priming, purge, leak control, stock identification, tube replacement and cleaning for the dosing subsystem.

Mandatory interlocks

ConditionAutomatic response
Invalid or stale sensorblock feedback correction
Reservoir below minimum levelblock concentrated stock dosing
Reservoir above maximum levelblock water addition
Calibration expiredblock volumetric automatic dosing
Maximum dose/runtime exceededstop channel and latch fault
Mixing delay activeblock second feedback action
Stock mismatchreject recipe execution
Cover/service state unsafedisable pump motion where required
Communication lossoutputs return to defined safe state
Leak detectedstop all liquid actuators and alert

Tube lifecycle

Tube service life is not published as a fixed 1,000-hour value unless validated for the exact tube, occlusion, fluid, speed and duty cycle. Replacement is triggered by calibrated-flow drift, visible damage, loss of occlusion, contamination risk or the validated service threshold.

Cleaning record

cleaning_id, system_id, channel_id, cleaning_agent,
concentration, contact_time, rinse_volume, verification_method,
operator, timestamp, next_allowed_fluid_class